Here, we can keep track of key measures from our two-photon microscope systems, in order to monitor for problems and understand the past history of the microscopes:

datebergpower at sample at 100% (milli Watts)Extinction ratio (power out of Pockel cell at 100% / power in at 0%)Pockel cell holding voltageAlignment assessment
04/27/2022III115537 (537/1)-0.52Workable. Re-alignment needed at low priority
12/16/2022IV117-0.29Quick alignment starting at the periscope because power at sample was 70mW
1/26/2023IV~150~800 (goes up to 1000)-0.31Much improved beam shape and power! Full realignment starting at the laser aperture, trhough half wave plate, P cell and allignment inside the scope. Fixed scanimage calibration issues to improved read out extinction ratio.Will add picture later, pretty circular